Product Name: Creatine kinase isoenzymes MB (Fluorescence Immunoassay)
Abbreviation: CK-MB
APPLICATION:
For the quantitative determination of creatine kinase isoenzyme MB in human serum, plasma or whole blood.
Creatine Kinase Isoenzymes MB (CK-MB) is one of the three isoenzymes of creatine kinase. CK-MB is mainly present in the myocardium, and other tissues are almost non-existent. Clinically Isozyme (CK-MB) is considered to be a specific and sensitive indicator of myocardial injury, and Its total activity ratio with CK can be used as a basis for diagnosis of acute myocardial infarction.
PRCINCIPLE:
Take the sample to the strip, The CK-MB in the specimen binds to the antibody coated with the fluorescent particles coated on the glass fiber to form a fluorescent particle-antibody-antigen complex. The immune complex is further chromatographed along the nitrocellulose membrane to the detection zone (T), and combined with the pre-coated mouse anti-CK-MB monoclonal antibody, the fluorescence intensity is proportional to the CK-MB content in the sample, and the remaining Fluorescent antibody particles were chromatographed into the control region (C) and bound to pre-coated goat anti-mouse IgG.
Method:
Fluorescence Immunoassay
Assay Time: 15 min
Sample Volume: 20ul
Sample Type: Serum / Plasma / Whole blood
Technical parameters:
Analytical Sensitivity: ≤1 ng/mL
Measuring Range: 1~100 ng/mL
Precision: CV≤15%
Stability: Stable up to 18 months if stored at 2 - 30°C in its sealed pouch.
STORAGE AND STABILITY:
The strip is stable for 18months (while sealed in an aluminum foil pouch) if stored at 2 ° C - 30 ° C. Do not freeze After the strip is opened, the test should be performed in 1 hour.
The date of manufacture is detailed in the label.
Package:
25 pieces/box
Name | Composition |
Strip | Consisting of a cartridge and a strip, the test strip comprises a sample pad, a glass fiber (coated with a fluorescent particle-CK-MB antibody conjugate), a nitrocellulose membrane (the detection zone is coated with a CK-MB monoclonal antibody, and the quality control zone is coated with a sheep) Anti-mouse IgG), absorbent paper, PVC board. |
Magcard | Have Calibration curve information of this batch reagents. |
Buffer | Main component is phosphate buffered saline (PBS). |
QC(Optional) | Freeze-dried products, the target value of each batch is shown in the target value list. |
Manual | To descripe basic information, usage methods and precautions |
Applicable POCT Analyzer
D-10, D-20, KHF01