Creatine Kinase Isoenzymes MB Determination Kit

Product Name: Creatine kinase isoenzymes MB  (Fluorescence Immunoassay

Abbreviation: CK-MB

 

APPLICATION:

For the quantitative determination of creatine kinase isoenzyme MB in human serum, plasma or whole blood.

Creatine Kinase Isoenzymes MB (CK-MB) is one of the three isoenzymes of creatine kinase. CK-MB is mainly present in the myocardium, and other tissues are almost non-existent. Clinically Isozyme (CK-MB) is considered to be a specific and sensitive indicator of myocardial injury, and Its total activity ratio with CK can be used as a basis for diagnosis of acute myocardial infarction.

 

PRCINCIPLE:

Take the sample to the strip, The CK-MB in the specimen binds to the antibody coated with the fluorescent particles coated on the glass fiber to form a fluorescent particle-antibody-antigen complex. The immune complex is further chromatographed along the nitrocellulose membrane to the detection zone (T), and combined with the pre-coated mouse anti-CK-MB monoclonal antibody, the fluorescence intensity is proportional to the CK-MB content in the sample, and the remaining Fluorescent antibody particles were chromatographed into the control region (C) and bound to pre-coated goat anti-mouse IgG.

 

Method:

Fluorescence Immunoassay

Assay Time: 15 min

Sample Volume: 20ul

Sample Type: Serum / Plasma / Whole blood

 

Technical parameters:

Analytical Sensitivity: ≤1 ng/mL

Measuring Range: 1~100 ng/mL

Precision: CV≤15%

Stability: Stable up to 18 months if stored at 2 - 30°C in its sealed pouch.

 

STORAGE AND STABILITY:

The strip is stable for 18months (while sealed in an aluminum foil pouch) if stored at 2 ° C - 30 ° C. Do not freeze After the strip is opened, the test should be performed in 1 hour.

The date of manufacture is detailed in the label.

 

Package:

25 pieces/box

Name

Composition

Strip

Consisting of a cartridge and a strip, the test strip comprises a sample pad, a glass fiber (coated with a fluorescent particle-CK-MB antibody conjugate), a nitrocellulose membrane (the detection zone is coated with a CK-MB monoclonal antibody, and the quality control zone is coated with a sheep) Anti-mouse IgG), absorbent paper, PVC board.

Magcard

Have Calibration curve information of this batch reagents.

Buffer

Main component is phosphate buffered saline (PBS).

QC(Optional)

Freeze-dried products, the target value of each batch is shown in the target value list.

Manual

To descripe basic information, usage methods and precautions

 

Applicable POCT Analyzer

D-10, D-20, KHF01