Product Name: Troponin I Determination Kit (Fluorescence Immunoassay)
Abbreviation: cTnI
APPLICATION:
For the quantitative detection of troponin I in human serum, plasma or whole blood.
PRCINCIPLE:
Cardiac troponin I (cTnI) is a 22.5KD myocardial protein. It forms a cardiac troponin complex with troponin T (TnT) and troponin C (TnC), to complete basic function of intracellular actin of calcium signaling between interactions. It has high specificity and diagnostic sensitivity in the diagnosis of acute myocardial infarction.
Method:
Fluorescence Immunoassay
Assay Time: 15 min
Sample Volume: 20µl
Sample Type: Serum / Plasma / Whole blood
Technical parameters:
Analytical Sensitivity: ≤0.1ng/ml
Measuring Range: 0.1 ~ 40 ng/mL
Precision: CV≤15%
Stability: Stable up to 18 months if stored at 2 - 30°C in its sealed pouch.
STORAGE AND STABILITY:
The strip is stable for 18months (while sealed in an aluminum foil pouch) if stored at 2 ° C - 30 ° C. Do not freeze After the strip is opened, the test should be performed in 1 hour.
The date of manufacture is detailed in the label.
Package:
25 pieces/box
Name | Composition |
Strip | Consisting of a cartridge and a strip, the test strip comprises a sample pad, a glass fiber (coated with a fluorescent particle- cTnI antibody conjugate), a nitrocellulose membrane (the detection zone is coated with a cTnI monoclonal antibody, and the quality control zone is coated with a sheep) Anti-mouse IgG), absorbent paper, PVC board. |
Magcard | Have Calibration curve information of this batch reagents. |
Buffer | Main component is phosphate buffered saline (PBS). |
QC(Optional) | Freeze-dried products, the target value of each batch is shown in the target value list. |
Manual | To descripe basic information, usage methods and precautions |
Applicable POCT Analyzer
D-10