Product Name: N-terminal pro-B-type natriuretic peptide (Fluorescence Immunoassay)
Abbreviation: NT-proBNP
APPLICATION:
For the quantification of N-terminal pro-brain natriuretic peptides in human serum, plasma or whole blood.
Used for auxiliary diagnosis of heart failure in clinic.
PRCINCIPLE:
Take the sample to the strip, The NT-proBNP in the specimen binds to the antibody coated with the fluorescent particles coated on the glass fiber to form a fluorescent particle-antibody-antigen complex. The immune complex is further chromatographed along the nitrocellulose membrane to the detection zone (T), and combined with the pre-coated mouse anti-NT-proBNP monoclonal antibody, the fluorescence intensity is proportional to the NT-proBNP content in the sample, and the remaining Fluorescent antibody particles were chromatographed into the control region (C) and bound to pre-coated goat anti-mouse IgG.
Method:
Fluorescence Immunoassay
Assay Time: 15 min
Sample Volume: 30µl
Sample Type: Serum / Plasma / Whole blood
Technical parameters:
Analytical Sensitivity: ≤ 50 pg/mL
Measuring Range: 50 ~ 20000 pg/mL
Precision: Intra-assay CV<10%, Inter-assay CV<15%.
STORAGE AND STABILITY:
The strip is stable for 18months (while sealed in an aluminum foil pouch) if stored at 2 ° C - 30 ° C. After the strip is opened, the test should be performed in 1 hour. The date of manufacture is detailed in the label.
Package:
25 pieces/box
Name | Composition |
Strip | Consisting of a cartridge and a strip, the test strip comprises a sample pad, a glass fiber (coated with a fluorescent particle- NT-proBNP antibody conjugate), a nitrocellulose membrane (the detection zone is coated with a NT-proBNP monoclonal antibody, and the quality control zone is coated with a sheep) Anti-mouse IgG), absorbent paper, PVC board. |
Magcard | Have Calibration curve information of this batch reagents. |
Buffer | Main component is phosphate buffered saline (PBS). |
QC(Optional) | Freeze-dried products, the target value of each batch is shown in the target value list. |
Manual | To descripe basic information, usage methods and precautions |
Applicable POCT Analyzer
D-10, D-20, KHF01