Product Name: Mycophenolic acid (Fluorescence Immunochromatography Assay)
Abbreviation: MPA
APPLICATION:
For the quantitative determination of mycophenolic acid in human serum, plasma or whole blood.
Mycophenolic acid is a commonly used immunosuppressant in organ transplantation. Free mycophenolic acid has pharmacological activity, and its metabolism is affected by many factors, resulting in individual differences in blood drug concentration. The content of enzymatic phenolic acid in human body is determined, and the dosage regimen is designed according to the individual characteristics and disease state of the patient, and provides a reference for achieving individualized administration, improving curative effect and reducing adverse reaction of the medicine clinically.
PRCINCIPLE:
Take the sample to the strip, The MPA in the specimen binds to an antibody coated with a fluorescent particle coated on the glass fiber to form a fluorescent particle-antibody-antigen complex. The immune complex is chromatographed along the nitrocellulose membrane to the detection zone (T), which binds to the pre-coated MPA-BSA conjugate and its fluorescence intensity is inversely proportional to the MPA content of the sample. The anti-IgY fluorescent antibody particles are chromatographed to the quality control zone (C) and bound to the pre-coated Organic IgY.
Method:
Fluorescence Immunoassay
Assay Time: 15 min
Sample Volume: 20µl
Sample Type: Serum / Plasma / Whole blood
STORAGE AND STABILITY:
The strip is stable for 18months (while sealed in an aluminum foil pouch) if stored at 2 ° C - 30 ° C, the fluorescent antibody in a tube is stable for 18months if stored at 2 ° C - 8 ° C, and the dissociation liquid in a tube is stable for 18months if stored at 2 ° C - 30 ° C. After the strip is opened, the test should be performed in 1 hour.
The date of manufacture is detailed in the label.
Package:
25 pieces/box
Name | Composition |
Strip | Consisting of a cartridge and a strip, the test strip comprises a sample pad, a glass fiber (Coated with fluorescent particles -MPA antibody conjugate, fluorescent particle - IgY antibody conjugate), nitrocellulose membrane (detection zone coated with MPA-BSA conjugate, quality control zone coated with organic IgY), absorbent paper, PVC board. |
Magcard | Have Calibration curve information of this batch reagents. |
Buffer | Main component is phosphate buffered saline (PBS). |
QC(Optional) | Freeze-dried products, the target value of each batch is shown in the target value list. |
Manual | To descripe basic information, usage methods and precautions |
Applicable POCT Analyzer
D-10, D-20, KHF01