Product Name: Digoxin (Fluorescence Immunoassay)
Abbreviation: DIG
APPLICATION:
For quantitative detection of digoxin in human serum, plasma or whole blood.
Digoxin is a cardiac glycoside drug, which is mainly used for the treatment of congestive heart failure, atrial fibrillation and therefore high atrial flutter. However, the therapeutic window of the drug is narrow and the individual difference is large. Even conventional doses may cause vomiting, nausea, palpitations, chest tightness and other poisoning phenomena. Therefore, it is a great significance to ensure the safety of clinical medication for the detection of digoxin blood concentration.
PRCINCIPLE:
Take the sample to the strip, The Digoxin in the specimen binds to an antibody coated with a fluorescent particle coated on the glass fiber to form a fluorescent particle-antibody-antigen complex. The immune complex is chromatographed along the nitrocellulose membrane to the detection zone (T), which binds to the pre-coated Digoxin-BSA conjugate and its fluorescence intensity is inversely proportional to the Digoxin content of the sample. The anti-IgY fluorescent antibody particles are chromatographed to the quality control zone (C) and bound to the pre-coated organic IgY.
Method:
Fluorescence Immunoassay
Assay Time: 15 min
Sample Volume: 40µl
Sample Type: Serum / Plasma / Whole blood
STORAGE AND STABILITY:
The strip is stable for 18months (while sealed in an aluminum foil pouch) if stored at 2 ° C - 30 ° C, the fluorescent antibody in a tube is stable for 18months if stored at 2 ° C - 8 ° C, and the dissociation liquid in a tube is stable for 18months if stored at 2 ° C - 30 ° C. After the strip is opened, the test should be performed in 1 hour.
The date of manufacture is detailed in the label.
Package:
25 pieces/box
Name | Composition |
Strip | Consisting of a cartridge and a strip, the test strip comprises a sample pad, a glass fiber (Coated with fluorescent particles-Digoxin antibody conjugate, fluorescent particle-IgY antibody conjugate), nitrocellulose membrane (detection zone coated with Digoxin-BSA conjugate, quality control zone coated with organic IgY), absorbent paper, PVC board. |
Magcard | Have Calibration curve information of this batch reagents. |
Buffer | Main component is phosphate buffered saline (PBS). |
QC(Optional) | Freeze-dried products, the target value of each batch is shown in the target value list. |
Manual | To descripe basic information, usage methods and precautions |
Applicable POCT Analyzer
D-10, D-20, KHF01